化工学报 ›› 2008, Vol. 59 ›› Issue (11): 2768-2773.

• 传递现象 • 上一篇    下一篇

含湿率对多孔介质热导率测量准确性的影响

于明志;隋晓凤;彭晓峰;方肇洪   

  1. 山东建筑大学热能工程学院;清华大学热能工程系
  • 出版日期:2008-11-05 发布日期:2008-11-05

Influence of moisture content on measurement accuracy of porous media thermal conductivity

YU Mingzhi;SUI Xiaofeng;PENG Xiaofeng;FANG Zhaohong   

  • Online:2008-11-05 Published:2008-11-05

摘要: 本文利用HotDisk热常数分析仪对不同含湿率时含湿沙的热导率进行测量研究,并对孔隙尺度水分形态、分布及演化过程进行了观察实验研究。研究发现,低含湿率(<25%)情况下,介质的有效热导率难以准确测得。低含湿时孔隙内水分主要以孤立液桥形式存在,测试过程中,贴近探头附近的水分由于受探头加热发生蒸发扩散而逐渐减少,导致探头周围试样构成持续发生变化,因而测量得到的热导率准确性较差。含湿率较高时,孔隙内水分相互联通,测试时水分蒸发扩散速度相对较慢,且存在毛细回流现象,探头周围试样构成维持恒定,因此可以准确获得热导率。

关键词:

多孔介质, 含湿率, 热导率, 测量

Abstract:

Thermal conductivity measurement accuracy of sands was experimentally studied with a hot disk thermal constant analyzer and water morphologies,distribution and evolution at pore scale were observed with charge coupled device(CCD)combined with microscope.It was found that thermal conductivities of samples with a low moisture content(<25%)could not be accurately measured.For samples with a low moisture content,the analysis showed that the water in the region adjacent to the analyser sensor mainly existed as isolated liquid bridges between/among sand particles and would evaporate and diffuse to relatively far regions because of being heated by the sensor during measurement.Water evaporation and diffusion caused the sample composition in the region adjacent to the sensor vary throughout the whole measurement process,and accordingly low accuracy of the obtained thermal conductivities.Due to high water connectivity in pores,the rate of water evaporation and diffusion in porous media of high moisture content was relatively slow compared with that in porous media of low moisture content.Meanwhile,water in the relatively far region flowed back to the region adjacent to the sensor by capillary force.Therefore,sample composition of the region adjacent to the sensor maintained constant and thermal conductivities of porous media with a relatively high moisture content could be measured with high accuracy.

Key words:

多孔介质, 含湿率, 热导率, 测量