Wet gas performance of two-path V turn reflection ultrasonic flow meter

ZHANG Qiang,XU Ying,ZHANG Tao   

  • Received:2012-01-06 Revised:2012-04-25 Online:2012-08-05 Published:2012-08-05
  • Supported by:

    National Natural Science Foundation of China

V形反射式双声道超声波流量计湿气测量特性

张强,徐英,张涛   

  1. 天津大学电气与自动化工程学院,天津市过程检测与控制重点实验室,天津 300072
  • 通讯作者: 徐英
  • 基金资助:

    国家自然科学基金

Abstract: In order to investigate the wet gas performance of ultrasonic flow meter,the wet gas flow characteristics were taken into account and the experiment was carried out with a two-path V turn reflection ultrasonic flow meter.The test result showed that the wet gas over-reading of ultrasonic flow meter was dependent on the L-M parameter.Liquid hold up and depth of liquid at the pipe bottom were the leading factors that affected the over-reading of ultrasonic flow meter with reflection sonar path.Consideration was given to liquid hold up,and a theoretical wet gas correlation of ultrasonic flow meter was deduced,the gas flow rates could be predicted within ±3% error under the flow condition when the L-M parameter was less than 0.1.This research work is useful for developing wet gas flow meter based on ultrasonic technology.

Key words: wet gas metering, ultrasonic flow meter, over-reading